色色网图片,亚洲 无码 制服 日韩 中文字幕,国产不卡免费在线视频,91久久亚洲国产成人精品性色

筆式PH計的操作說明

2020-09-24 16:16:02 1519


在實驗室中,對于一些精度要求不高的情況,我們可以考慮使用筆式PH計筆式PH計是采用很新微功耗積體電路設計,性能優越,廣泛應用于理化分析,環保,食品,化妝品,水族館,養殖業,無土栽培…等等行業。

ph計,酸度計


優點:

 技術性能:

·全防水,IP67等級。

·測量范圍:0~14pH

·可更換電極。

·精確度:±0.1pH

·超大熒幕顯示。

·環境溫度:0℃~50℃

·攜帶輕便。

·電池:紐扣電池1.5V×3


·儀器尺寸:170mm×37mm×37mm


· 儀器凈重:80g

儀器校準:打開PH筆頂蓋,調校調節器

1.       將電極放入PH7.0的緩沖液中,調校右邊調節器,使儀器顯示為7.0。

2.       將電極放入清水中清洗。

3.       將電極放入PH4.0的緩沖液中,調校左邊調節器,使儀器顯示為4.0。

維護:

l          PH測試筆在出廠時已校準,可直接使用。(長期久置不用,請于使用時重新校準)

l          測量后,電極應用干凈水清洗。

l          當儀器顯示屏變暗時,請更換電池及注意電池正負極方向。

使用:

1.       將電極保護蓋取下,打開電源開關。

2.       將電極放入測試液體中,儀器便會顯示準確PH讀數。

3.       測量完畢,關閉電源,套上電極保護蓋。

PH PEN/Tester Operation Manual

PH PEN/Tester adopted with latest micro-electronic circuit design, to provide high performance, steady and accuracy measurement. This PEN is widely applies in work of chemical analysis, environmental protection, food, hygiene, aquaculture and horticulture, etc.

Advantage

 Specification

·Waterproof Class IP67

·Testing Range: 0.0pH to 14.0pH

·Replaceable Electrode

·Accuracy: ±0.1pH

·Large LCD display

·Ambient temperature:0℃ to 50℃

·Extremely portable

·Battery: LR44×3


·Dimension: 158mm×30mm×30mm


· Weight : 67g

Instrument Calibration

1.          Soak the electrode in 7.0pH/4.0pH solution. If the display does not show the right pH value, adjust the unit to the right measurement. (Remove the protection cap and adjust the knob. Right knob for 7.0pH / Left knob for 4.0pH).

2.          Clean the electrode with distilled water.

3.          Soak the electrode into buffer solution at 4.0pH. If the display does not show the right pH value then adjust the left point to make it displays 4.0pH.

4.          If it is necessary please repeat procedure 1. to 3.

Maintenance

l          If the instrument has not been operating for a long time, it should be recalibrated before use.

l          Usually clean the electrode with distilled water after use.

l          Replace the batteries when the display became dim.

Operating

1.          Remove the protection cap and turn on the power.

2.          Place the electrode in sample solution until the result show on the display.

3.          Turn off the power, clean the electrode and cover the protection cap after test.

更多關于筆式PH計的知識請聯系合肥卓爾儀器儀表有限公司的技術人員。

標簽: 筆式PH計